Our goal is to create an accurate, affordable, and portable device to better characterize muscle spasticity. Current prevalent methods involving the Modified Ashworth Scale leave something to be desired. By using a streamlined quantitative methodology and developing a streamlined smartphone application, clinicians all around the world are now one step closer to a more accurate way of determining different degrees of muscle spasticity in patients.
|
Our device measures three critical components of spasticity, namely, range of motion, velocity, and resistance to motion. With these three key parameters, the torque can be calculated at a range of different velocities. This data then translates directly to a numerical value defining spasticity. The interface is intuitive and the method requires no background knowledge, thus making it extremely simple for any physician, nurse, or therapist to implement for patient use.
|
Preliminary Presentation:
Progress Presentation:
Final Presentation:
|
|
|
Design Safe Analysis (.pdf) | |
File Size: | 22 kb |
File Type: |
Characteristic |
Original Specification |
Outcome |
---|---|---|
Cost |
< $200 |
$74.39 |
Portability |
Lasts at least 10 hours |
Lasts 27 hr |
Accuracy of FSR |
< ±1N |
±0.2N |
Accuracy of Accelerometer |
< ±0.05g |
±0.0153g |
Accuracy of Velocity |
< 5% |
< 2.5% |
Reliability of FSR |
< 10% between trials |
2% between trials |
Reliability of Accelerometer |
< 10% between trials |
0.28% between trials |
Size |
< 21.6cm x 19cm x 5cm (average size of a physician’s coat pocket) |
13.5cm x 8cm x 8cm and can be disassembled |
Weight |
< 0.5 kg |
0.432 kg |
Training |
< 10 minutes |
< 10 minutes |
Versatility |
Accommodate limb size of 15 cm to 45 cm in circumference. Measure force from 0N to 200N. |
Accommodates limb size of 13cm to 35cm in circumference. Measures forces from 0N to 200N. |
Durability |
Withstand 200N in compression and 30N in tension |
Withstands > 200N in compression and 60.48N in tension |
Ease of Use |
Duration < 10 minutes |
Depends on # of tests, but likely < 10 min |
Frequency |
> 10 Hz |
10 Hz |
Power Consumption |
< 1 W |
0.25W |
Range |
< 5 m away |
10 m away |
Memory |
< 2 kilobytes |
2 kilobytes |
Bandwidth |
< 2 Mbps |
2 Mbps |